1. Introdução
This manual provides comprehensive instructions for the safe and efficient operation of your U.S. Solid Precision Lab Scale. Please read this manual thoroughly before use and retain it for future reference. This analytical balance is designed for precise weighing in laboratory, industrial, and educational settings, offering a capacity of 5000g with 0.01g readability.
2. Precauções de segurança
- Always operate the scale on a stable, level, and vibration-free surface.
- Evite expor a balança a temperaturas extremas, umidade ou luz solar direta.
- Não sobrecarregue a balança além de sua capacidade máxima de 5000g.
- Keep the weighing pan clean and free from debris.
- Utilize somente o adaptador de energia fornecido.
- Avoid sudden impacts or dropping the scale.
3. Configuração e instalação
3.1 Desembalagem
Carefully remove the scale and all accessories from the packaging. Verify that all components listed in the packing list are present. The package should include the scale, stainless steel weighing pan, and power adapter.
3.2 Posicionamento
Place the scale on a firm, level surface away from drafts, vibrations, and direct heat sources. Ensure adequate space around the scale for operation and ventilation.
3.3 Nivelando a Balança
The scale must be perfectly level for accurate measurements. Use the adjustable feet at the bottom of the scale and the bubble level indicator to achieve proper leveling.

Imagem: Frente view of the U.S. Solid Precision Lab Scale, showing the bubble level indicator on the left side of the display panel.

Image: Diagram illustrating how to adjust the leveling feet until the bubble is centered, ensuring proper horizontal adjustment for accuracy.
3.4 Power Connection and Warm-up
Connect the provided power adapter to the scale and then to a suitable power outlet. For optimal accuracy, allow the scale to warm up for at least 30 minutes after connecting to power or after a power loss before performing any measurements.
Video: This video demonstrates the initial setup of the U.S. Solid Precision Lab Scale, including connecting the power supply and the recommended 30-minute warm-up period for optimal performance.
4. Instruções de operação
4.1 Pesagem Básica
- Pressione o LIGADO/DESLIGADO botão para ligar a escala.
- Wait for the display to show "0.00g".
- Place the item to be weighed on the center of the stainless steel pan.
- Leia o peso estável exibido na tela LCD.
4.2 Conversão de Unidades
The scale supports 19 measurement units. Press the UNIDADE button to cycle through the available units (g, kg, ct, T, TAR, dr, PKT, GN, TMR, gsm, tlJ, mo, dwt, oz, lb, tlT, ozt, tlH, %).

Image: Visual representation of the 19 different measurement units supported by the U.S. Solid Precision Lab Scale, including grams, kilograms, carats, and more.
4.3 Função Tara
To subtract the weight of a container, place the empty container on the pan and press the TARE/CAL button. The display will reset to "0.00g". You can then add the item to be weighed into the container, and only the net weight of the item will be displayed.
4.4 Função de contagem de peças
The scale can count pieces of uniform weight. To use this function:
- Coloque comoample quantity (e.g., 10, 20, 50, 100, 150, 200, 250, or 500 pieces) of the items to be counted on the pan.
- Pressione e segure o COU Pressione o botão até que "PCS" apareça no visor.
- Use o UNIDADE botão para selecionar o sample quantity that matches the number of items on the pan.
- Imprensa TARE/CAL to confirm. The display will now show the number of pieces.
- Adicione mais itens à panela e a balança exibirá a contagem total.

Image: Illustration of the piece counting feature, demonstrating how to set sample sizes (10, 20, 50, etc.) for accurate item counting.
4.5 Configuração da luz de fundo
To adjust the backlight behavior:
- Pressione o COU button to choose between C0-OFF, C0-1, or C0-5.
- C0-OFF: Backlight is always on.
- C0-1: Backlight will turn off after 1 second of inactivity.
- C0-5: Backlight will turn off after 5 seconds of inactivity.
- Por fim, clique TARA until the program ends and returns to 0.00.
Video: This segment demonstrates how to adjust the backlight settings of the scale, allowing users to choose between always-on, 1-second, or 5-second inactivity timeouts for the backlight.
4.6 Unit Package Locking
To lock specific unit packages for quick access:
- Pressione e segure o COU button. Release it when "C0-OFF" is displayed.
- Clique TARA to choose C06-1.
- Pressione o COU button to cycle through C06-1, C06-2, up to C06-16, each representing a different combination of units.
- Por fim, clique TARA until the program ends and returns to 0.00.
Video: This segment illustrates how to lock specific unit packages on the scale, allowing users to customize and quickly switch between preferred sets of measurement units.
5. RS232 Interface Communication
The U.S. Solid Precision Lab Scale features an RS232 interface for connecting to a computer or printer, enabling serial data exchange. This allows for data logging and integration with other systems.

Imagem: Lado view of the U.S. Solid Precision Lab Scale, highlighting the RS232 interface port for external connectivity.
5.1 Key Output Mode Setting (C5-2)
In Key Output Mode, the scale sends weighing data to the connected device only when the UNIDADE botão é pressionado.
- Pressione e segure o COU button. Release it when "C0-OFF" is displayed.
- Clique TARA to choose C5-0.
- Pressione o COU button to choose C5-2.
- Por fim, pressione o botão TARA button to enter into key output mode.
- To output data, open "Uartassist" software (or similar serial terminal software), set Serial Port to COM3, Baud Rate to 9600, Data Bits to 8, Parity to None, and Stop Bits to One. Click "Open serial port".
- Cada vez que você pressiona o UNIDADE button on the scale, the computer will output the current weighing data.
Video: This segment demonstrates how to set the scale to "Key Output" mode (C5-2) for USB interface testing. It shows connecting the scale to a computer and manually triggering data output by pressing the UNIT button.
5.2 Continuous Output Mode Setting (C5-0)
In Continuous Output Mode, the scale continuously sends weighing data to the connected device without requiring manual input.
- Pressione e segure o COU button. Release it when "C0-OFF" is displayed.
- Clique TARA to choose C5-0.
- Por fim, pressione o botão TARA button to enter into continuous output mode.
- With "Uartassist" software (or similar serial terminal software) open and configured (COM3, 9600 baud, 8 data bits, no parity, one stop bit), click "Open serial port". The weighing results will be continuously output on the computer.
Video: This segment demonstrates how to set the scale to "Continuous Output" mode (C5-0) for USB interface testing. It shows the scale continuously sending weighing data to a connected computer.
6. Calibração
Regular calibration ensures the accuracy of your scale. The scale supports manual calibration.
- Ensure the scale is on a stable surface and has completed its warm-up period.
- With the scale displaying "0.00g", press and hold the TARE/CAL button until "CAL" appears.
- The display will then show a flashing weight value (e.g., "2000g"). This is the required calibration weight.
- Carefully place the specified calibration weight onto the center of the pan.
- Wait for the display to stabilize and show "PASS" or return to "0.00g".
- Remova o peso de calibração. A balança agora está calibrada.

Image: Diagram illustrating the calibration process, emphasizing the need to recalibrate after moving the balance or before use, and to hold the TARE/CAL button for instructions.
7. Manutenção e Cuidados
- Limpeza: Wipe the scale's exterior with a soft, damp cloth. Do not use abrasive cleaners or immerse the scale in water. The stainless steel pan can be removed for separate cleaning.
- Armazenar: When not in use for extended periods, store the scale in a dry, dust-free environment.
- Inspeção: Verifique periodicamente se o cabo de alimentação apresenta algum dano.
8. Solução De Problemas
| Problema | Possível causa | Solução |
|---|---|---|
| Balança não liga | No power, faulty adapter, or power outlet issue. | Check power connection, try a different outlet, ensure adapter is functional. |
| Leituras imprecisas | Not level, drafts, vibrations, not warmed up, needs calibration, or overloaded. | Level the scale, move to a stable location, allow warm-up, perform calibration, ensure weight is within capacity. |
| O visor mostra "O-Ld" | Sobrecarga. | Remova imediatamente o excesso de peso para evitar danos. |
| Falha de comunicação RS232 | Incorrect port settings, cable issue, or wrong output mode. | Verify COM port, baud rate, data bits, parity, and stop bits. Check cable connection. Ensure correct output mode (Key or Continuous) is selected. |
9. Especificações
- Marca: Sólido dos EUA
- Modelo: 5000g x 0.01g Analytical Balance
- Capacidade: 5000 gramas (5 kg)
- Legibilidade: 0.01 Grams (10 mg)
- Tipo de exibição: LCD com luz de fundo
- Interface: RS232
- Unidades de medida: 19 units (g, kg, ct, T, TAR, dr, PKT, GN, TMR, gsm, tlJ, mo, dwt, oz, lb, tlT, ozt, tlH, %)
- Aplicações: Check-weighing, Tare, Piece Counting, Manual Calibration, Overload Protection
- Material: Carcaça ABS e panela de aço inoxidável
- Dimensões do produto: 8.9 x 8.7 x 3.2 polegadas
- Peso do item: 2.9 libras
- Poder: Adaptador AC (incluído)
10. Garantia e Suporte
For warranty information and technical support, please refer to the warranty card included with your product or visit the official U.S. Solid webGuarde o comprovante de compra como prova de aquisição.
Informações de contato: Consulte o manual oficial do fabricante. website or contact details provided in the product packaging for support.