The Extech PH90 Waterproof pH Meter is a precision instrument designed for accurate pH measurement across various applications, including water quality testing, hydroponics, aquaculture, swimming pools, beverage production, and laboratory use. This device features a robust waterproof housing, a replaceable electrode, 3-point automatic calibration, and auto buffer recognition for ease of use and reliable results. Its compact design makes it suitable for both field and professional environments.

Extech PH90 Waterproof pH Meter

Figure 1: Extech PH90 Waterproof pH Meter. This image shows the compact, handheld pH meter with its digital display and control buttons, ready for use.

2. Principais características

3. Configuração e preparação

3.1. Instalação da bateria

The PH90 meter requires two CR2032 Lithium Metal batteries (included). To install or replace batteries:

  1. Unscrew the battery compartment cap located at the top of the meter.
  2. Insert the two CR2032 batteries, ensuring correct polarity (+/-).
  3. Securely screw the battery compartment cap back into place.

3.2. Electrode Preparation

Before first use, and after storage, prepare the electrode:

  1. Remove the protective cap from the electrode tip. The cap contains a moist sponge to keep the electrode hydrated.
  2. Rinse the electrode tip thoroughly with distilled or deionized water. Do not wipe the electrode, as this can create static charges.
  3. Allow any excess water to drip off.

3.3. Calibração inicial

For optimal accuracy, perform a 3-point calibration before initial use and periodically thereafter. Calibration buffers (pH 4, 7, and 10) are sold separately.

  1. Ligue o medidor pressionando o LIGADO/DESLIGADO botão.
  2. Immerse the electrode into a pH 7.00 buffer solution.
  3. Pressione e segure o CAL button until "CAL" appears on the display. The meter will automatically recognize the buffer and calibrate.
  4. Enxágue o eletrodo com água destilada.
  5. Repeat the process with pH 4.00 and pH 10.00 buffer solutions. The meter will automatically detect and calibrate to these points.
  6. Assim que a calibração estiver concluída, o medidor retornará ao modo de medição.
Extech PH90 Meter Display and Buttons

Figure 2: Close-up of the Extech PH90 display and control buttons, including ON/OFF, MODE, and CAL buttons.

4. Instruções de operação

4.1. Fazendo uma medição

  1. Certifique-se de que o medidor esteja calibrado.
  2. Rinse the electrode with distilled water and gently shake off excess.
  3. Mergulhe a ponta do eletrodo no sampa solução, garantindo que o sensor esteja totalmente submerso.
  4. Gently stir the solution to ensure homogeneity and allow the reading to stabilize. This may take up to a minute, especially if the sample temperature differs significantly from the meter's previous environment.
  5. Read the pH value displayed on the LCD screen. The temperature will also be displayed.
  6. After measurement, rinse the electrode thoroughly with distilled water.

4.2. Função Hold

Para congelar a leitura atual no visor, pressione o SEGURAR button (if available, typically integrated with MODE or a separate button). Press it again to release the hold and return to live measurement.

4.3. Compensação Automática de Temperatura (ATC)

The Extech PH90 features Automatic Temperature Compensation, which adjusts pH readings based on the sample's temperature, ensuring accuracy across varying temperatures. The temperature is displayed alongside the pH value.

5. Manutenção

5.1. Cuidados e armazenamento dos eletrodos

Proper electrode care is crucial for maintaining accuracy and extending the life of your meter:

Extech PH90 Electrode Tip

Figure 3: Close-up of the Extech PH90's replaceable electrode tip, which requires proper care and storage.

5.2. Substituição do eletrodo

The PH90 features a replaceable electrode. If the electrode performance degrades significantly despite proper cleaning and calibration, it may need replacement. Refer to the Extech webPara obter informações sobre a compra, visite o site ou entre em contato com o suporte ao cliente.asing replacement electrodes (Part PH95).

5.3. Substituição da bateria

When the battery indicator appears on the display or the meter fails to power on, replace the two CR2032 Lithium Metal batteries as described in Section 3.1.

6. Solução De Problemas

ProblemaPossível causaSolução
Leituras imprecisas ou instáveis
  • Electrode not calibrated
  • Dirty or dry electrode
  • Temperature not equilibrated
  • Soluções tampão vencidas
  • Eletrodo danificado
  • Perform a 3-point calibration.
  • Clean the electrode and ensure it is properly stored.
  • Allow sufficient time for the meter's temperature sensor to equilibrate with the sampeu.
  • Use fresh calibration buffer solutions.
  • Substitua o eletrodo.
O medidor não liga.
  • Baterias descarregadas
  • Instalação incorreta da bateria
  • Replace batteries with new CR2032 cells.
  • Verifique a polaridade da bateria.
"CAL" error during calibration
  • Solução tampão incorreta usada
  • Contaminated buffer solution
  • Electrode issue
  • Ensure correct pH buffer (4, 7, or 10) is used.
  • Use fresh, uncontaminated buffer solutions.
  • Clean or replace the electrode.

7. Especificações

ParâmetroValor
Número do modeloPH90
Faixa de pH0 a 14 pH
Precisão de pH±0.01 pH
Faixa de temperatura(Not explicitly stated, but integrated for ATC)
Fonte de energia2 x CR2032 Lithium Metal batteries
Peso do item2.88 onças (aproximadamente 81.6 gramas)
Dimensões do produto6.7 x 1.4 x 1.4 polegadas (aproximadamente 17 x 3.6 x 3.6 cm)
Componentes incluídosFlat surface pH electrode, cap, batteries (pH buffers sold separately)
FabricanteEx-tecnologia

8. Garantia e Suporte

For warranty information, technical support, or to purchase replacement parts and accessories, please visit the official Extech website or contact Extech customer service. You can also find more information and products at the Loja Extech na Amazon.

Documentos relacionados - PH90

Préview Guia do Usuário dos Medidores de pH à Prova d'Água Extech ExStik™ PH100 ​​e PH110
Guia completo do usuário para os medidores de pH à prova d'água Extech ExStik™, modelos PH100 ​​e PH110. Saiba mais sobre operação, calibração, manutenção e especificações.
Préview Manual do Usuário do Medidor de pH de Palma à Prova d'Água Extech PH220
Manual do usuário do medidor de pH portátil à prova d'água Extech PH220. Abrange introdução, descrição do medidor, operação, calibração, medição, armazenamento, substituição da bateria, manutenção, solução de problemas e especificações.
Préview Extech MO257 Pinless Moisture Meter User Guide
User guide for the Extech MO257 Pinless Moisture Meter, detailing its features, operation, specifications, and maintenance for accurate non-invasive moisture measurements in building materials.
Préview Manual do Usuário do Testador de Continuidade Extech ET40 para Serviço Pesado
Manual do usuário do testador de continuidade Extech ET40 para serviço pesado. Aprenda como testar com segurança e eficácia a continuidade de componentes desenergizados, fusíveis, interruptores, relés, fiação e placas de circuito.
Préview Extech Electromagnetic Field Meter Model 480823 User Manual
User manual for the Extech Electromagnetic Field Meter, Model 480823. This document details the meter's specifications, operation, battery replacement, EMF exposure information, warranty, and calibration services. It measures EMF in Gauss and Tesla units with a frequency bandwidth of 30 to 300Hz.
Préview Manual do Usuário dos Medidores de pH à Prova d'Água Extech ExStik PH100 ​​e PH110
Manual do usuário completo para os medidores de pH à prova d'água Extech ExStik, modelos PH100 ​​e PH110. Fornece instruções detalhadas sobre operação, calibração, manutenção dos eletrodos, substituição da bateria, precauções de segurança, especificações e solução de problemas para esses dispositivos de teste de pH de alta precisão.