1. Introdução
The AMTAST WA-60A Portable Water Activity Meter is designed for precise measurement of water activity (aw) in various substances, including food products like beef jerky, freeze-dried food, fruits, and vegetables. This device helps assess product preservation and stability by quantifying the free water content. Its portable design and integrated sensor make it a convenient tool for on-site analysis.
2. Produto acabadoview

Figura 2.1: The AMTAST WA-60A Water Activity Meter in use, displaying a water activity reading. This device measures free water activity to assess the preservation and stability of food, medicine, and other products.

Figura 2.2: The AMTAST WA-60A Water Activity Meter with its sample plate. The kit typically includes the meter, sample plate, portable bag, and manual for accurate measurement of free water content.
Principais características:
- Design portátil: Weighing approximately 100 grams, this meter is lightweight and suitable for on-the-go testing.
- Medidas precisas: Offers a measurement range of 0-1.0aw with a resolution of ±0.01aw and an accuracy of 0.02aw.
- Sensor Integrado: Features a non-conductive humidity sensor and integrated readings in a single unit, simplifying water activity analysis.
- Teste rápido: Provides rapid results, typically completing measurements within 5 minutes.
- Tela clara: Equipped with a 6-digit backlit LCD screen that simultaneously displays water activity (aw) and temperature.
- Funções fáceis de usar: Includes automatic and manual shut-off options, along with a low battery indicator.

Figura 2.3: Detalhado view of the meter's backlit LCD screen, showing both the water activity (aw) and temperature values for clear readings.

Figura 2.4: Image showing the battery compartment for 2x1.5V AAA batteries and the non-conductive humidity sensor. The device also features automatic shutdown, manual shutdown, and alarm indication.
3. Especificações

Figura 3.1: The AMTAST WA-60A Water Activity Meter shown alongside its key specifications.
| Parâmetro | Valor |
|---|---|
| Número do modelo | WA-60A |
| Faixa de medição | 0.00 ~ 1.00 aw |
| Resolução | ±0.01 aw |
| Precisão | ±0.02 aw |
| Temperature Operation Conditions | 0-50°C |
| Humidity Operation Conditions | <95% UR |
| Peso do item | 100 gramas (3.53 onças) |
| Fonte de energia | 2 x 1.5V AAA Size (UM-4) Battery (not included) |
| Material | abdômen |
| Dimensões | Approximately 13.5 cm (5.3 inches) height, 7 cm (2.8 inches) width |
4. Configuração
Before first use, ensure the device is properly set up.
- Instalação da bateria: Open the battery compartment on the back of the meter. Insert two 1.5V AAA batteries, ensuring correct polarity. Close the compartment securely.
- Sampo Titular: The meter comes with a sample holder featuring a strong magnetic connection. Ensure the sample holder is clean and dry before use.
- Calibration (Highly Recommended): For accurate and reliable results, especially for food safety applications, it is highly recommended to calibrate the meter using certified water activity standards. This should be done in the range relevant to your testing.
4.1 vídeo: This video demonstrates the setup and basic operation of the AMTAST WA-60A Water Activity Meter, including how to prepare samples for measurement.
5. Operação
Follow these steps for accurate water activity measurement:
- Preparar Sampem: Coloque seus sample into a clean, dry sample cup. Ensure the sample fills the cup adequately without overflowing.
- Locaisampem: Posicione o sample cup onto the base of the meter. The magnetic sample holder will secure it in place.
- Equilíbrio: Passo crucial: After placing the sample and closing the meter, allow sufficient time for the sample to equilibrate with the sensor. This process can take several hours, or even overnight, depending on the sample type and environmental conditions. Do not turn on the meter immediately.
- Ligar: Uma vez queample has equilibrated, turn on the meter while it is still closed on the sampeu.
- Leitura de Medição: The LCD screen will display the water activity (aw) and temperature. Allow the reading to stabilize before recording the final value.
- Desligar: The device features automatic and manual shut-off options. To manually turn off, press and hold the power button.
5.1 vídeo: Este vídeo oferece uma visão geral.view of the AMTAST Smart Water Activity Meter, demonstrating its use in a laboratory setting.
5.2 vídeo: This video demonstrates the process of using the Water Activity Meter for monitoring food water activity (AW).

Figura 5.1: The meter is designed for quick and efficient testing, providing accurate readings for water activity measurements.

Figura 5.2: The AMTAST WA-60A is widely used for testing water activity in various food products, including dried fruits, beef jerky, cookies, and sausages.
6. Manutenção
- Limpeza: Mantenha o sample holder and sensor area clean and free from debris. Use a soft, dry cloth to wipe the device. Do not use abrasive cleaners or immerse the device in water.
- Armazenar: Store the meter in its protective case in a cool, dry place when not in use. Remove batteries if storing for extended periods to prevent leakage.
- Calibração: Regular calibration with known standards is essential to maintain accuracy. Refer to Section 4 for calibration recommendations.
7. Solução De Problemas
| Problema | Possível causa | Solução |
|---|---|---|
| Leituras imprecisas ou erráticas | Insufficient equilibration time; lack of calibration; dirty sensor. | Garantirample equilibrates for several hours before measurement. Calibrate the device regularly with appropriate standards. Clean the sensor and samptitular. |
| O dispositivo não liga | Baterias descarregadas ou instaladas incorretamente. | Check battery polarity. Replace batteries with fresh 1.5V AAA batteries. |
| A tela está escura ou ilegível | Bateria fraca. | Substitua as pilhas. |
8. Informações de segurança
- Não tente desmontar ou reparar o dispositivo por conta própria. Contate pessoal qualificado para assistência técnica.
- Mantenha o dispositivo longe de temperaturas extremas, umidade e luz solar direta.
- Manuseie o dispositivo com cuidado para evitar danos físicos.
- Descarte as baterias de acordo com os regulamentos locais.
9. Garantia e Suporte
For warranty information, technical support, or service inquiries, please contact AMTAST customer service. Refer to the product packaging or the official AMTAST website para detalhes de contato.
AMTAST Official Store: Visit the AMTAST Store on Amazon