UNI-T UPO1102

UNI-T UPO1102 100MHz 2-Channel Digital Oscilloscope Instruction Manual

Model: UPO1102

1. Introdução

This instruction manual provides essential information for the safe and effective use of the UNI-T UPO1102 100MHz 2-Channel Digital Oscilloscope. The UPO1102 is a versatile instrument designed for electronic and electrical measurements, featuring a 100MHz bandwidth, 1 GSa/s real-time sample rate, 56 Mpts memory depth, and a 7-inch display. It incorporates Ultra Phosphor technology for enhanced waveform display and analysis capabilities.

Leia atentamente este manual antes de utilizar o dispositivo para garantir o seu correto funcionamento e evitar danos.

2. Informações de segurança

Observe sempre as seguintes precauções de segurança para evitar ferimentos pessoais e danos ao instrumento ou a outros dispositivos conectados.

3. Produto acabadoview

3.1 Painel Frontal

Frente view of the UNI-T UPO1102 Digital Oscilloscope, showing the 7-inch display, control knobs, and input connectors.

Figure 3.1: Front Panel of UPO1102 Oscilloscope

The front panel features the main display, various control knobs and buttons for waveform adjustment, trigger settings, and menu navigation. Input channels CH1 and CH2 are located at the bottom right.

3.2 Painel traseiro

Traseira view of the UNI-T UPO1102 Digital Oscilloscope, showing the power input, USB ports, and ventilation.

Figure 3.2: Rear Panel of UPO1102 Oscilloscope

The rear panel includes the AC power input, USB ports for data transfer and printing, and ventilation grilles to ensure proper cooling during operation.

3.3 Dimensões

Dimensions of the UNI-T UPO1102 Digital Oscilloscope, showing a width of 13.23 inches, height of 6.46 inches, and depth of 4.13 inches.

Figure 3.3: UPO1102 Oscilloscope Dimensions

The physical dimensions of the UPO1102 are approximately 13.23 inches (width) x 6.46 inches (height) x 4.13 inches (depth).

4. Configuração

4.1 Desembalagem e Inspeção

Upon receiving your UPO1102 oscilloscope, carefully unpack all components and inspect them for any signs of damage. Retain the packaging for future transport or storage.

Acessórios padrão:

4.2 Conexão de Energia

  1. Ensure the oscilloscope is placed on a stable, level surface with adequate ventilation.
  2. Conecte o cabo de alimentação fornecido à entrada de alimentação CA no painel traseiro do osciloscópio.
  3. Conecte a outra extremidade do cabo de alimentação a uma tomada elétrica CA aterrada.
  4. Pressione o botão liga/desliga no painel frontal para ligar o osciloscópio.

4.3 Conexão e Compensação da Sonda

Before taking measurements, it is crucial to compensate the probes to ensure accurate readings.

  1. Connect a probe to one of the input channels (CH1 or CH2) on the front panel.
  2. Connect the probe tip to the "Probe Comp" terminal and the ground clip to the ground terminal next to it.
  3. Pressione o Auto button. The oscilloscope will automatically adjust the settings to display the compensation signal.
  4. Observe the square wave displayed on the screen. If the waveform is not a perfect square (over-compensated or under-compensated), adjust the trimmer on the probe body using a non-metallic tool until a flat-top square wave is achieved.
  5. Repeat for the second probe if using both channels.
Example of a square wave displayed on the oscilloscope screen during probe compensation.

Figure 4.1: Probe Compensation Waveform

A correctly compensated probe displays a clean square wave. Adjust the probe's trimmer if the waveform shows overshoot or undershoot.

5. Instruções de operação

5.1 Controles Básicos

5.2 Waveform Acquisition and Display

The UPO1102 features a 7-inch display and Ultra Phosphor technology for clear waveform visualization, even with complex signals.

Oscilloscope display showing a complex waveform with varying intensity, demonstrating Ultra Phosphor technology.

Figure 5.1: Ultra Phosphor Display

Ultra Phosphor technology provides a gradient display, where brighter areas indicate more frequent occurrences of the waveform, helping to visualize signal anomalies and variations.

Oscilloscope display showing two overlaid waveforms, demonstrating multi-channel measurement.

Figure 5.2: Multi-Channel Waveform Display

The UPO1102 can display and analyze two channels simultaneously, allowing for comparison of different signals.

5.3 Funções de Medição

5.4 Digital Decoding

The innovative hardware decoding feature allows for real-time decoding of serial bus protocols. With a deep storage of 56 Mpts, decoding speed is maintained at millisecond levels, preventing delays when viewing decoded data. This function does not affect the waveform refresh rate, maintaining the digital fluorescence display effect.

Oscilloscope display showing decoded serial data alongside the raw waveform.

Figure 5.4: Digital Decoding Example

The decoding feature displays protocol information directly on the screen, simplifying debugging of serial communications.

5.5 Data Storage and Connectivity

6. Manutenção

6.1 Cuidados Gerais

6.2 Limpeza

Para limpar a parte externa do osciloscópio:

  1. Disconnect the power cord and all probes/cables.
  2. Use um pano macioampened with mild detergent and water. Do not use abrasive cleaners or solvents.
  3. For the display, use a soft, lint-free cloth specifically designed for electronic screens.
  4. Ensure the instrument is completely dry before reconnecting power.

7. Solução De Problemas

This section addresses common issues you might encounter with the UPO1102 oscilloscope. For problems not listed here, contact UNI-T customer support.

ProblemaPossível causaSolução
Sem exibição após ligar.Power cord not connected, power outlet faulty, instrument fault.Check power cord connection. Test power outlet. If problem persists, contact support.
Nenhuma forma de onda exibida.Probe not connected, signal too small/large, trigger not set correctly, channel off.Ensure probe is connected to signal source. Press Auto button. Adjust vertical scale (Volts/Div) and horizontal scale (Sec/Div). Check trigger level. Ensure channel is enabled.
Forma de onda instável.Trigger level incorrect, trigger mode unsuitable, signal noise.Adjust trigger level. Try different trigger modes (e.g., Normal or Single). Reduce noise in the signal path.
Incorrect measurements.Probe compensation incorrect, probe attenuation setting wrong, measurement settings incorrect.Perform probe compensation (Section 4.3). Verify probe attenuation setting matches physical probe (e.g., 1X, 10X). Check measurement parameters.

8. Especificações

ParâmetroValor
Largura de banda100 MHz
Canais2
S em tempo realample Taxa1 GSa/s
Profundidade de memória56 Mpts
Taxa de captura de forma de onda500,000 wfm/s
Mostrar7 polegadas
FFT SampPontos de espera1 Mpts
Dimensões do produto13 x 4 x 6 polegadas (aproximadamente 330 x 102 x 152 mm)
Peso do item8.13 libras (aproximadamente 3.69 kg)
FabricanteUNIDADE
ASINB0D7F2D9MS
Data da primeira disponibilidade19 de junho de 2024

9. Garantia e Suporte

For warranty information, technical support, or service inquiries regarding your UNI-T UPO1102 oscilloscope, please refer to the official UNI-T website ou entre em contato com o departamento de atendimento ao cliente.

Loja oficial da UNI-T: Visit the UNI-T Store on Amazon

Please have your product model (UPO1102) and ASIN (B0D7F2D9MS) available when contacting support.

Documentos relacionados - UPO1102

Préview Manual do Usuário da Sonda Ativa de Terminação Única UNI-T UT-PA2000/UT-PA1000
Este manual do usuário fornece informações completas sobre as sondas ativas de terminação única UNI-T UT-PA2000 e UT-PA1000, abrangendo suas características, especificações, diretrizes de segurança, operação, manutenção e garantia.
Préview Guia Rápido do Osciloscópio Digital de Fósforo Série UNI-T UPO1000
Um guia HTML conciso e otimizado para SEO para o osciloscópio digital de fósforo da série UNI-T UPO1000, abordando segurança, configuração, operação e solução de problemas. Inclui descrições detalhadas dos painéis frontal e traseiro e dos elementos da interface do usuário.
Préview Manual do Usuário do Osciloscópio de Fósforo Ultra Série UNI-T UPO2000E
Manual do usuário completo para o osciloscópio de fósforo Ultra da série UNI-T UPO2000E, abrangendo operação, segurança, recursos e especificações para modelos como UPO2072E, UPO2102E, UPO2074E e UPO2104E.
Préview MSO7000X Series Mixed Signal Oscilloscope User's Manual
User's manual for the MSO7000X Series Mixed Signal Oscilloscope, detailing its features, operation, safety precautions, and technical specifications. Covers setup, vertical and horizontal systems, trigger system, protocol decoding, measurements, mathematical operations, and troubleshooting.
Préview Folha de dados dos osciloscópios de alta resolução da série UNI-T MSO1000HD
Ficha técnica completa dos osciloscópios de alta resolução da série UNI-T MSO1000HD, detalhando recursos, especificações, características de desempenho, pontas de prova, acessórios e informações para pedidos.
Préview Folha de dados do osciloscópio de sinal misto da série UNI-T MSO3000X
Ficha técnica do osciloscópio de sinal misto da série UNI-T MSO3000X, detalhando suas características, especificações e capacidades, incluindo sua largura de banda de 500 MHz e taxa de amostragem de 5 GSa/s.amptaxa de atualização e interface de tela sensível ao toque de 10.1 polegadas.